분동세척관련 OIML기술기준

 

 

 Cleaning weights

 

 

 

 

 

It is important to clean weights before any measurements are made because the cleaning process may change the mass of the weight. Cleaning should not remove any significant amounts of weight material. Weights should be handled and stored in such a way that they stay clean. Before calibration, dust and any foreign particles shall be removed. Care must be taken not to change the surface properties of the weight (i.e. by scratching the weight). 

 

 

If a weight contains significant amounts of dirt that cannot be removed by the methods cited above, the weight or some part of it can be washed with clean alcohol, distilled water or other solvents. Weights with internal cavities should normally not be immersed in the solvent to avoid the possibility that the fluid will penetrate the opening. If there is a need to monitor the stability of a weight in use, the mass of the weight should, if possible, be determined before cleaning.

 

 

 

After weights are cleaned with solvents they must be stabilized for the times given in Table B.1.

 

 

 

Table B.1 Stabilization time after cleaning

 

 

 

Weight class

E1

E2

F1

F2 to M3

After cleaning with alcohol

7–10 days

3–6 days

1–2 days

1 hour

After cleaning with distilled water

4–6 days

2–3 days

1 day

1 hour

 

 

 

 

 

Thermal stabilization

 

 

Prior to performing any calibration tests, the weights need to be acclimated to the ambient conditions of the labo- ratory. In particular, weights of classes E1, E2 and F1 should be close to the temperature in the weighing area.

 

B.4.3.1 The mandatory minimum times required for temperature stabilization (depending on weight size, weight class and on the difference between the initial temperature of the weights and the room temperature in the labora- tory) are shown in Table B.2. As a practical guideline, a waiting time of 24 hours is recommended.

 

 

 

Table B.2 [11] Thermal stabilization in hours

 

 

 

ÄT*

Nominal value

Class E1

Class E2

Class F1

Class F2

 

 

 

± 20 °C

1 000, 2 000, 5 000 kg

-

-

79

5

100, 200, 500 kg

-

70

33

4

10, 20, 50 kg

45

27

12

3

1, 2, 5 kg

18

12

6

2

100, 200, 500 g

8

5

3

1

10, 20, 50 g

2

2

1

1

< 10 g

1

0.5

 

 

 

± 5 °C

1 000, 2 000, 5 000 kg

-

-

1

1

100, 200, 500 kg

-

40

2

1

10, 20, 50 kg

36

18

4

1

1, 2, 5 kg

15

8

3

1

100, 200, 500 g

6

4

2

0.5

10, 20, 50 g

2

1

1

0.5

< 10 g

0.5

 

 

 

www.seoho.biz

서호 전북 전주시 덕진구 혁신로 586 (54851)

Tel. 063-214-9325 / Fax. 063-214-9328

seoho@seoho.biz

 

 


  1. No Image 14Jun
    by
    Views 8812 

    기체유량단위

  2. No Image 12Feb
    by
    Views 4882 

    부품의 극성 및 양부 판별법

  3. No Image 12Feb
    by
    Views 4656 

    테스터의 기본 사용법

  4. No Image 19Mar
    by
    Views 4112 

    압력계의 분류

  5. 열전대의 원리와 종류

  6. No Image 19Mar
    by
    Views 3111 

    배관ANSI규격

  7. No Image 19Mar
    by
    Views 2500 

    열전대 전압 대 온도 도표

  8. No Image 14Jun
    by
    Views 2292 

    SI유도단위의 유도량, 명칭, 기호

  9. No Image 27Mar
    by
    Views 2210 

    유량계(면적식)

  10. No Image 19Apr
    by
    Views 1823 

    방폭의 일반

  11. 온도센서 Thermocouple 과 RTD 를 통한 정확한 온도 측정 – 진동의 간섭을 방지하는 법

  12. No Image 19Mar
    by
    Views 1349 

    관용 나사의 종류

  13. No Image 30May
    by 서호
    Views 1252 

    분동세척관련 OIML기술기준

목록
Board Pagination Prev 1 2 3 4 Next
/ 4
위로