분동세척관련 OIML기술기준

 

 

 Cleaning weights

 

 

 

 

 

It is important to clean weights before any measurements are made because the cleaning process may change the mass of the weight. Cleaning should not remove any significant amounts of weight material. Weights should be handled and stored in such a way that they stay clean. Before calibration, dust and any foreign particles shall be removed. Care must be taken not to change the surface properties of the weight (i.e. by scratching the weight). 

 

 

If a weight contains significant amounts of dirt that cannot be removed by the methods cited above, the weight or some part of it can be washed with clean alcohol, distilled water or other solvents. Weights with internal cavities should normally not be immersed in the solvent to avoid the possibility that the fluid will penetrate the opening. If there is a need to monitor the stability of a weight in use, the mass of the weight should, if possible, be determined before cleaning.

 

 

 

After weights are cleaned with solvents they must be stabilized for the times given in Table B.1.

 

 

 

Table B.1 Stabilization time after cleaning

 

 

 

Weight class

E1

E2

F1

F2 to M3

After cleaning with alcohol

7–10 days

3–6 days

1–2 days

1 hour

After cleaning with distilled water

4–6 days

2–3 days

1 day

1 hour

 

 

 

 

 

Thermal stabilization

 

 

Prior to performing any calibration tests, the weights need to be acclimated to the ambient conditions of the labo- ratory. In particular, weights of classes E1, E2 and F1 should be close to the temperature in the weighing area.

 

B.4.3.1 The mandatory minimum times required for temperature stabilization (depending on weight size, weight class and on the difference between the initial temperature of the weights and the room temperature in the labora- tory) are shown in Table B.2. As a practical guideline, a waiting time of 24 hours is recommended.

 

 

 

Table B.2 [11] Thermal stabilization in hours

 

 

 

ÄT*

Nominal value

Class E1

Class E2

Class F1

Class F2

 

 

 

± 20 °C

1 000, 2 000, 5 000 kg

-

-

79

5

100, 200, 500 kg

-

70

33

4

10, 20, 50 kg

45

27

12

3

1, 2, 5 kg

18

12

6

2

100, 200, 500 g

8

5

3

1

10, 20, 50 g

2

2

1

1

< 10 g

1

0.5

 

 

 

± 5 °C

1 000, 2 000, 5 000 kg

-

-

1

1

100, 200, 500 kg

-

40

2

1

10, 20, 50 kg

36

18

4

1

1, 2, 5 kg

15

8

3

1

100, 200, 500 g

6

4

2

0.5

10, 20, 50 g

2

1

1

0.5

< 10 g

0.5

 

 

 

www.seoho.biz

서호 전북 전주시 덕진구 혁신로 586 (54851)

Tel. 063-214-9325 / Fax. 063-214-9328

seoho@seoho.biz

 

 


List of Articles
번호 제목 최근 수정일 조회 수
32 국가표준기본법 [시행 2018. 12. 13] [법률 제15643호, 2018. 6. 12, 일부개정] file 2021.04.16 248
31 금속표준용어집(2009) file 2018.03.20 278
30 기체유량단위 2019.06.14 9078
29 방폭의 일반 2018.08.07 1863
28 배관ANSI규격 2018.08.07 3223
27 배관JIS규격 2018.08.07 549
26 부품의 극성 및 양부 판별법 2019.02.12 4994
25 분동 허용 오차 file 2019.02.12 731
» 분동세척관련 OIML기술기준 file 2018.08.07 1265
23 분동의 선택방법 file 2021.02.22 407
22 압력 측정 방법 2018.08.07 326
21 압력계(액주식 ) file 2018.12.14 958
20 압력계(탄성식) 2018.12.14 596
목록
Board Pagination Prev 1 2 3 4 Next
/ 4
위로