분동세척관련 OIML기술기준

 

 

 Cleaning weights

 

 

 

 

 

It is important to clean weights before any measurements are made because the cleaning process may change the mass of the weight. Cleaning should not remove any significant amounts of weight material. Weights should be handled and stored in such a way that they stay clean. Before calibration, dust and any foreign particles shall be removed. Care must be taken not to change the surface properties of the weight (i.e. by scratching the weight). 

 

 

If a weight contains significant amounts of dirt that cannot be removed by the methods cited above, the weight or some part of it can be washed with clean alcohol, distilled water or other solvents. Weights with internal cavities should normally not be immersed in the solvent to avoid the possibility that the fluid will penetrate the opening. If there is a need to monitor the stability of a weight in use, the mass of the weight should, if possible, be determined before cleaning.

 

 

 

After weights are cleaned with solvents they must be stabilized for the times given in Table B.1.

 

 

 

Table B.1 Stabilization time after cleaning

 

 

 

Weight class

E1

E2

F1

F2 to M3

After cleaning with alcohol

7–10 days

3–6 days

1–2 days

1 hour

After cleaning with distilled water

4–6 days

2–3 days

1 day

1 hour

 

 

 

 

 

Thermal stabilization

 

 

Prior to performing any calibration tests, the weights need to be acclimated to the ambient conditions of the labo- ratory. In particular, weights of classes E1, E2 and F1 should be close to the temperature in the weighing area.

 

B.4.3.1 The mandatory minimum times required for temperature stabilization (depending on weight size, weight class and on the difference between the initial temperature of the weights and the room temperature in the labora- tory) are shown in Table B.2. As a practical guideline, a waiting time of 24 hours is recommended.

 

 

 

Table B.2 [11] Thermal stabilization in hours

 

 

 

ÄT*

Nominal value

Class E1

Class E2

Class F1

Class F2

 

 

 

± 20 °C

1 000, 2 000, 5 000 kg

-

-

79

5

100, 200, 500 kg

-

70

33

4

10, 20, 50 kg

45

27

12

3

1, 2, 5 kg

18

12

6

2

100, 200, 500 g

8

5

3

1

10, 20, 50 g

2

2

1

1

< 10 g

1

0.5

 

 

 

± 5 °C

1 000, 2 000, 5 000 kg

-

-

1

1

100, 200, 500 kg

-

40

2

1

10, 20, 50 kg

36

18

4

1

1, 2, 5 kg

15

8

3

1

100, 200, 500 g

6

4

2

0.5

10, 20, 50 g

2

1

1

0.5

< 10 g

0.5

 

 

 

www.seoho.biz

서호 전북 전주시 덕진구 혁신로 586 (54851)

Tel. 063-214-9325 / Fax. 063-214-9328

seoho@seoho.biz

 

 


List of Articles
번호 제목 최근 수정일 조회 수
44 기체유량단위 2019.06.14 8773
43 부품의 극성 및 양부 판별법 2019.02.12 4867
42 테스터의 기본 사용법 2019.07.24 4645
41 압력계의 분류 2018.08.07 4109
40 열전대의 원리와 종류 1 2018.08.07 3653
39 배관ANSI규격 2018.08.07 3100
38 열전대 전압 대 온도 도표 2018.08.07 2492
37 SI유도단위의 유도량, 명칭, 기호 2019.06.14 2285
36 유량계(면적식) 2018.12.14 2210
35 방폭의 일반 2018.08.07 1821
34 온도센서 Thermocouple 과 RTD 를 통한 정확한 온도 측정 – 진동의 간섭을 방지하는 법 2018.12.14 1815
33 관용 나사의 종류 2018.08.07 1343
» 분동세척관련 OIML기술기준 file 2018.08.07 1252
목록
Board Pagination Prev 1 2 3 4 Next
/ 4
위로